Bundle-imaging Fiber photometry Optogenetic Collimator
-
This collimator brings optogenetic illumination to the Bundle-imaging setup. Its collimated beam diameter depends strongly on the light source divergence, fiber NA and collimator focal length. For example, lasers are typically injected into 0.22 NA optical fibers with a 0.2 NA lens that yields an effective 0.20 NA at the other end of the fiber. On the other hand, Ce:YAG light source is highly divergent and should be used with 0.5 NA optical fibers and corresponding collimators.
The beam diameter after collimation determines the size of the illuminated circle on the fiber bundle end face. Therefore, when selecting this collimator, the user must choose appropriate light sources and illuminated circle sizes that match his bundle type. If the user changes the fiber bundle type for some reason, he should change the collimators accordingly.Number & diameter of fibers in a bundle
Bundle diameter (mm) Laser illuminated circle diameter
(mm)Ce:YAG illuminated circle diameter
(mm)4x 200 µm 0.56 0.75 0.90 7x 200 µm 0.60 0.75 0.90 12x 200 µm 0.95 1.10 1.30 19x 200 µm 1.12 1.25 1.30 4x 400 µm 1.05 1.25 1.30 7x 400 µm 1.20 1.40 1.30 12x 400 µm 1.75 1.90 1.90 19x 400 µm 2.10 2.20 3.00 -
Wavelength range 350 to 700 nm Field of view selected by table in DESCRIPTION section BFMC objective 10x - NA 0.40 Optical fiber connector FC/PC
Type
Source
Effective NACollimator
Focal Length
( mm )Illuminated diameter
( mm )'AD'
achromatic
doublet
0.20 12.0
0.75
17.5 1.10 20.0 1.25 22.5 1.40 30.0 1.90 'AH'
aspheric
high NA
0.50 3.0 0.90 4.3 1.30 5.5 1.60 8.0 1.90 12.5 3.00
*valid with BFMC objective 10x - NA0.40.
-
Drawing Fluorescence Mini Cube (FC connector on all ports)
Brochure Fluorescence Mini Cube (FC connector on all ports)
- le choix d'une sélection entraîne une actualisation complète de la page